Product Info
The ALTAIR 4X Multigas Detector is for use by trained and qualified personnel. It is designed to be used when performing a hazard assessment to assess potential worker exposure to combustible and toxic gases and vapors as well as low level of oxygen, and to determine the appropriate gas and vapor monitoring needed for a workplace.
Preventive Maintenance Schedule
Next interval:
- Perform a full calibration periodically to ensure accuracy and immediately if the instrument fails the Calibration Check
Daily:
- Perform a Calibration Check before each day’s use to verify proper instrument operation
As needed:
- Perform a Calibration Check more frequently if the unit is subjected to physical shock or high levels of contaminants
- Perform a Calibration Check more frequently if the tested atmosphere contains materials that may desensitize the combustible gas sensor
Parts List
- Stainless Steel Suspender Clip: Clip for attaching the instrument to clothing (Part Number: 10069894)
- 34L Quad Gas Mix: Calibration gas mixture (2.5% CH4, 15% O2, 60 ppm CO) (Part Number: 10049056)
- 34L Quad Gas Mix: Calibration gas mixture (2.5% CH4, 15% O2, 60 ppm CO, 10 ppm NO2) (Part Number: 10058172)
- 58L Quad Gas Mix: Calibration gas mixture (2.5% CH4, 15% O2, 60 ppm CO, 10 ppm NO2) (Part Number: 10058171)
- Calibration Assembly: Includes calibration cap, tubing, and connector (Part Number: 10089321)
- North American Power Supply with Charge Connector: Power supply for charging the instrument (Part Number: 10087913)
- Charging Cradle Assembly with Power Supply (North America): Charging cradle and power supply for the instrument (Part Number: 10087368)
- Vehicle Charging Cradle Assembly: Charging cradle for use in vehicles (Part Number: 10095774)
- MSA Link Software CD-Rom: Software for configuring and managing the instrument (Part Number: 10088099)
- JetEye IR Adapter with USB Connector: Adapter for connecting the instrument to a PC (Part Number: 10082834)
- Combustible Sensor Replacement Kit: Replacement kit for the combustible gas sensor (Part Number: 10121212)
- O2 Sensor Replacement Kit: Replacement kit for the oxygen sensor (Part Number: 10106729)
- CO/H2S Two Toxic Sensor Replacement Kit: Replacement kit for the carbon monoxide and hydrogen sulfide sensors (Part Number: 10106725)
- CO/NO2 Two Toxic Sensor Replacement Kit: Replacement kit for the carbon monoxide and nitrogen dioxide sensors (Part Number: 10121217)
- Front Housing with Integrated Dust Filters (charcoal): Replacement front housing with charcoal dust filters (Part Number: 10110030)
- Front Housing with Integrated Dust Filters (Phosphorescent): Replacement front housing with phosphorescent dust filters (Part Number: 10110029)
- Main Board w/ Battery Pack: Replacement main board and battery pack (Part Number: 10124956)
- LCD Frame Assembly: Replacement LCD frame, LCD, zebra strips, and screws (Part Number: 10110061)
- Sensors Gasket, Screws (4x), self tapping (2x): Replacement sensor gasket and hardware (Part Number: 10110062)
- CD Manual ALTAIR 4X: Instruction manual on CD-ROM (Part Number: 10106623)
Product Specs Sheet
- Weight: 8.3 oz. (instrument with battery and clip)
- Dimensions: 4.4 x 3.00 x 1.37 inches – without fastening clip
- Alarms: Four gas alarm LEDs, a charge status LED, an audible alarm, and a vibrating alarm
- Volume of Audible Alarm: 95 dB typical
- Display: LCD display
- Battery Type: Rechargeable Li ION battery
- Instrument Run Time: 34 hours at 77°F (25°C)
- Charging Time: ≤ 4 hours
- Temperature Range: 14 to 104°F (-10 to 40°C) Normal operating range, -4 to 122°F (-20 to 50°C) Extended operating range, 50 to 95°F (10 to 35°C) While charging battery
- Humidity Range: 15 – 90% relative humidity, non-condensing; 5 – 95% RH intermittent
- Atmospheric Pressure Range: 800 to 1200 mbar
- Ingress Protection: IP 67
- Measuring Methods: Combustible gas: Catalytic sensor, O2 and Toxic gas: Electrochemical sensor
- Measuring Range: Combustible 0-100% LEL, 0-5.00% CH4, O2 0-30% Vol., CO 0-1999 ppm, H2S 0-200 ppm, NO2 0-50.0 ppm
Troubleshooting
- ERROR TEMP – Temperature is below -40°F (-40°C) or above 167°F (75°C):
- Return to normal temperature range and recalibrate
- Contact MSA
- ERROR EE – EEPROM Memory error:
- Contact MSA
- ERROR PRG – Flash Memory error:
- Contact MSA
- ERROR RAM – RAM Memory error:
- Contact MSA
- ERROR UNK – Unknown error:
- Contact MSA
- LOW BATTERY – Battery warning repeats every 60 seconds:
- Remove from service as soon as possible and recharge battery
- BATTERY ALARM – Battery is completely discharged:
- Instrument is no longer sensing gas. Remove from service and recharge battery
- ERROR CHARGE – Charge error:
- Instrument must be between 10 and 35°C to charge
- Contact MSA if problem persists
- SENSOR ERROR – Missing Sensor:
- Verify sensor is properly installed and recalibrate
- Contact MSA if problem persists
- Instrument does not turn ON:
- Low battery – Charge instrument
- Sensor warning – Sensor is near its end of life
- Sensor alarm – Sensor has reached end of life and cannot be calibrated, replace sensor and recalibrate
Please
note that any procedure, checklist, or other document available on Limble is provided for general education and
information only and does not constitute legal, medical, or financial advice. Limble makes the materials available
AS IS and AS PROVIDED, without warranties of any kind. By downloading or using any such materials, you assume the
risk that they may not be appropriate for your specific situation and agree that you are solely responsible for any
such use, including compliance with applicable law and with meeting any conditions of product warranties. It is
recommended to contact OEM for the most accurate information.
6fb2e4b0-40cd-4c73-872d-62eab252074a
Configure and custumize to fit your needs
-
Increase Productivity
Prioritize and complete work orders faster with efficient, collaborative software your team will love using.
-
Reduce Downtime
Avoid setbacks with a proactive, preventive maintenance strategy and move fast when emergencies do happen.
-
World Class Support
Limble’s support team is ready to answer any questions you have in 60 seconds or less, 24/7.
Want to see Limble in Action?
FAQs
Why are Standard Operating Procedures (SOP) Important?
SOPs ensure consistency, efficiency, and safety by providing clear guidelines for tasks. They streamline training, maintain quality, ensure compliance, and support continuous improvement and accountability.
What is CMMS?
CMMS stands for Computerized Maintenance Management System. CMMS software helps businesses manage, automate, and streamline all of their maintenance operations.
Is Limble’s CMMS Mobile Friendly?
Limble is consistently rated Easiest-to-Use CMMS on review sites like G2, Capterra, and Software Advice. And our customers agree. With our mobile CMMS app, teams experience 30%+ better productivity, on average, requiring little to no training or ramp-up time. Our CMMS app can travel with your team, no matter where they go! Visit our App Store or Google Play for more information.
Integration
Limble provides seamless, pre-built CMMS Integrations with the most widely used software systems. That means no help from a developer or your IT team is required. Learn more about our integrations.
Security
At Limble, our world-class data security practices ensure your account information is safe. We use state-of-the-art technologies and industry best practices to maintain a secure infrastructure, including SOC-II Type II certification, regular penetration testing, and continuous security training for our staff.